S.no POSTED ID DOM MODEL TYPE DESCRIPTION MANUFACTURER IMAGE
0-2004 Fonar Upright MRI Scanner Magnet Room Minimum Installed Area 504 sq. feet [28' x 18'] (46.8 sq. meters) Ceiling Height Required 11 feet (3.35 meters) Magnet Specification Weight 245,000 lbs. (111,130kg) which equals = 122.5 US tons (111.1 Metric tons) Assembled in pieces on-site Frame Outer Dimensions 102 in. (259cm) W x 107 in.(272 cm) H x 176 in. (447 cm) D Power Requirements 480 Volt, 3-phase, 125 Amps Power Consumption 100 KVA, The magnet is turned off when not in use Magnet Room Heat Load: 30,000 BTU Equipment Room Minimum Installed Area 240 sq. feet [12' x 20'] (21 sq. meters) Heat Load 48,000 BTU System Racks Power Requirements: 120/208 Volt, 3-phase, 200 Amps Power Consumption: 20 KVA Closed-loop air-cooled water chiller (included with system) Capacity 30 tons Power Requirements 480 Volt, 3-phase, 100 Amps Power Consumption 47 KVA (when running) The coolant is chilled in a coolant tank and repeatedly passed through a heat load where it absorbs heat from industrial processes or equipment. The heat absorbed from the heat load is then dissipated in an evaporator before returning to the coolant tank for re-cooling and recycling through the closed loop. The temperature of the coolant is regulated by a thermostat that automatically detects changes in the temperature within the coolant tank. This thermal controller then adjusts the refrigeration effect in the evaporator accordingly to ensure the fluid in the coolant tank is at an optimal temperature for its next interaction with the heat load It does have a Chiller but water based Located in MN 10-11-23 58215 2004 Fonar Upright MRI Fonar Fonar Fonar